3
General Instructions
1. Place the OxyGo FIT® in a well ventilated location.
Air intake and exhaust must have clear access.
Locate the OxyGo FIT® in such a way that any
auditory alerts may be heard. Always operate the
OxyGo FIT® in an upright position (see image for
proper orientation).
WARNING Avoid use of the OxyGo FIT® in presence of pollutants, smoke or fumes.
Do not use the OxyGo FIT® in presence of flammable anesthetics,
C
!
cleaning agents or other chemical vapors.
CAUTION
Do not obstruct air intake or exhaust when operating the device.
Blockage of air circulation or proximity to a heat source may lead to
internal heat buildup and shutdown or damage to the concentrator.
CAUTION
The OxyGo FIT® Concentrator is designed for continuous use. For optimal
sieve bed (columns) life, the product should be used frequently.
2. Ensure particle filter is in place.
CAUTION
Do not operate the OxyGo FIT® without the particle
filter in place. Particles drawn into the system may
damage the equipment.
Operating Instructions
Intake
Exhaust
Operating Instructions Chapter 3
This side
UP
Exhaust