OxyGo FIT® Column Change Procedure (continued)
9. Insert column assembly into the OxyGo FIT® concen-
trator. Do not leave the column ends exposed; col-
umn assembly should be inserted into the OxyGo FIT
as soon as the dust caps have been removed.
10. P ush the column assembly into the device such that
the columns are fully seated into the OxyGo FIT®
concentrator. The spring loaded latch button should
fully return to the closed position.
11. Connect the AC power supply cord to the OxyGo FIT®
and plug the power supply AC cord into an
electrical outlet. Do Not Power on the OxyGo FIT®
concentrator.
12. Press and hold the plus (+) and (-) minus buttons for
5 seconds. The screen will display "sieve reset"
message. Release button once message is displayed
on screen.
13. Press the alert
button once and screen will
display "sieve reset success".
14. P ress the power button to turn on the OxyGo
FIT®, and use normally.
15. If you are using My OxyGo App, navigate to the Advanced screen,
then to Additional Information screen and click on the Column Reset button.
Closed and locked
Cleaning, Care and Maintenance Chapter 6
37