Contains Transmitter Module IC: 12246A-BM71S2 Contains FCC ID: A8TBM71S2
This device complies with Part 15 of the FCC rules. Operation is subject to the following
two conditions: (1) this device may not cause harmful interference, and (2) this device
must accept any interference received, including interference that may cause undesired
operation.
Note: IT-network is a system composed of wireless (Bluetooth) transmission between the
OxyGo FIT® and the My OxyGo Application.
• Connection of the OxyGo FIT® to an IT-Network could result in previously unidentified
risks to patients, operators or third parties.
• Subsequent changes to the IT-network could introduce new risks and require
additional analysis.
• Changes to the IT-network include:
º Changes in the IT-network configuration.
º Connection of additional items to the IT-network.
º Disconnecting items from the IT-network.
º Updating equipment connected to the IT-network.
Classifications
Mode of Operation:
Type of Protection Against Electrical
Shock:
Degree of Protection to Concentrator
Components Against Electrical Shock:
Degree of Protection to Concentrator
Components Against Ingress of Water:
Degree of Safety for Application in
Presence of Anesthetic Gases:
*Normal position of the OxyGo FIT® is upright with the user interface display facing upward.
42
Continuous Duty
Class II
Type BF
Not intended for cardiac application
IP22 - Vertically dripping water shall have
no harmful effect & protect against
ingress of solid objects ≥ 12.5 mm.
diameter when the enclosure is tilted
at an angle up to 15˚ from its normal
position*
Not suitable for such application