Idiomas disponibles

5.
The PRESS START message will appear showing that measurement can
be started. Press the START+ switch. Depending on the position of the
TEST CURRENT/ control the LOW CURRENT message will turn up.
6.
Turning the TEST CURRENT/ control clockwise increases the current.
Adjust this control until obtaining the desired value or until the current indi-
cator (bar-graph) indicates 100%. The bar-graph will show the test current
value as a percentage of the nominal value selected by using the CUR-
RENT SELECTOR0.
7.
The lowest current for measuring is 10 % of the scale nominal value. It is
important to consider that the measurement errors increase while test cur-
rent decreases. The error is specified for test currents higher than 80%.
8.
The DISPLAY* will show the resistance value measured and the corre-
sponding unit (Ω [ohms], mΩ [milli-ohms] or µΩ [micro-ohms]).
9.
That value can be retained in the display by pressing the HOLD- key.
Pressing this key again, the value will be erased.
10. Press the STOP, switch in order to finish the measurement.
11. Finally, when finishing measurements, turn the equipment off using the
ON/OFF# switch.
Precaution: Do not connect or disconnect the test leads during the measurement.

Messages

CIRCUTOR MH-10
When turning the equipment on using the ON/OFF# switch, this introductory
message appears for a while. During that time, the equipment carries out some
functional checking.