Setting Up the Unit
1
1. Screw in the clamp screw.
2. Remove the battery cover with a #1 Phillips screwdriver.
3. Remove the battery pack from the console.
Note If replacing the battery pack, disconnect the battery wire connectors,
and properly discard the old battery (p. 12).
4. Connect battery wire connectors to battery pack.
Note Do not use force when aligning the latch to the tab.
5
5. Insert the battery pack into the console to the right of the battery wires.
6. Insert the battery cover and screw into place.
7. Position the console below the irrigation bag, 12" above the highest
surface of the patient. Tighten the clamp screw to mount the console to
the IV pole.
8. Connect the power cord to the console. Plug the cord into an electrical
outlet to charge the batteries
Caution
Ensure the unit is free from all liquid before plugging in.
Note The battery pack takes 1.5 hours to charge. "FC" will appear on the
flow rate display when fully charged.
8
2
6
3
7
4
8