WelchAllyn CONNEX 6000-ECG Instrucciones De Uso página 39

Módulo de ecg de dispositivos
Tabla de contenido
Instrucciones de uso
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the 901106 is used exceeds the applicable RF
compliance level above, the 901106 should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the 901106.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications
equipment and the 901106
The 901106 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or user of the 901106 can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the 901106 as recommended below,
according to the maximum output power of the communications equipment.
Rated max. output
power of transmitter
(W)
0.01
0.1
1
10
Electromagnetic immunity
3 Vrms
3 Vrms, 6 Vrms ISM
150 kHz to 80 MHz, 6
Vrms ISM freq.
3 V/m 80 MHz to 2.5
3 V/m
GHz
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
80 MHz to 800 MHz
d= (1.17)
d= (1.17)
0.12
0.12
0.37
0.37
1.17
1.17
3.70
3.70
Guía y declaraciones del fabricante sobre EMC 35
d= (1.17)
d= (1.17)
80 to 800 MHz
d= (2.33)
800 MHz to 2,5 GHz
whereP is the maximum output power rating of
the transmitter in watts (W) andd is the
recommended separation distance in meters
(m).
Interference may occur in the vicinity of
equipment marked with the following symbol:
800 MHz to 2.5 GHz
d= (2.23)
0.23
0.74
2.33
7.38
Tabla de contenido