Warning
• When the battery level is depleted i.e when the remaining charge indicator is low the actual load may be to high for the device to reach the
preset torque limit value. In this case (in the case of operation by batteries), the auto reverse function will not be activated. When high torque
is required, use this product in a condition where the batteries are not signifi cantly coansumed (As indicated by the battery indicator)
• If a load is continuously applied to the motor handpiece, it may automatically stop to prevent overheating. (The torque limit value indicates
"OK" on the LCD panel). In this case, leave the motor handpiece for a while until it cools down. When the display returns to normal, the motor
handpiece can be used.

9. ERROR AND REMEDY

ERROR MARK
REMEDY
• During performing the operation, in the case that the actual load
exceeds the torque value set, it will be automatically reversed. At
that time, when it exceeds the set torque it will be automatically
stopped and Error code (E-1) will be displayed on the screen.
10. MAINTENANCE OF THE PRODUCT
1. Cleaning
Angle Handpiece
• Rotate the head in a tap water about 10 seconds with maximum
speed to remove the remains (blood, physiologic saline solution,
etc.) and rinse.
• Do not put the tap water in the motor.
• Disconnect the angle handpiece to the motor handpiece.
• Remove contaminants on the angle handpiece using a nylon brush
after soaking for 20 minutes in Enzymatic Detergent.
• Rinse the angle handpiece using the tap water for 3 minutes to
remove Enzymatic Detergent.
• Wash the angle handpiece for 10 minutes in the ultrasonic cleaner
containing Enzymatic Detergent.
• Rinse the angle handpiece for 3 minutes in the ultrasonic cleaner
containing purifi ed water repeatedly subjected 3 times. (Change the
purifi ed water each time.)
• Wipe using dust-absorbent after the product soaked in alcohol(70%)
for 30 seconds.
12
BAE370R_IFU-Final.indd 12
• Remedy : Push Motor ON/OFF Button
Motor Handpiece
• Wipe the surface of motor handpiece with alcohol swab or paper
towel after removing the dirt.
• Pay attention to not enter the water into the motor handpiece.