Replacing the fuses
If the fuses have blown (see page 8, Fig. 1/10), proceed as follows to replace them:
Switch the autoclave off at the power switch and pull the power plug out from the socket.
1.
Open the door manually as described in the section "Door emergency open in case of power failure". Unscrew
2.
the two screw caps of the fuse holder (page 8, Fig. 1/10) on the bottom front of the autoclave using a
screwdriver or a coin.
On the inside of the door there are two replacement fuses (see marking).
Fig. 9 Replacement fuses on the inside of the door
Remove the faulty fuses and insert the new replacement fuses firmly into the fuse holder.
3.
Fig. 10 View front, bottom right
Screw the caps of the fuse holder on the bottom front of the autoclave on again.
4.
Insert the power plug of the autoclave back into the socket and switch the autoclave on again at the power
5.
switch.
If the fuses fail repeatedly, please contact the customer service of your specialist supplier.