Medela BiliBed Manual De Instrucciones página 12

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Hour meters
IX
1)
Hour meters
Maintenance
X
12
EN
1
Warning
Disconnect power supply before performing maintenance jobs! An annual routine
control of the displays on the hour meters and the function of the BiliBed is
recommended.
10.1 Changing the fluorescent tube
In a 1500 hour-interval, the light tube needs changing. By using the "total h" hour
meter this time interval can be monitored accurately.
The BiliBed is equipped with a "Light Tube Change Log"-card on the inside of the
light unit. On this log you will find a table as follows:
Recommended hours
Empfohlene Stunden
Nb. heures recom.
Duracion recomendada
The BiliBed is equipped with two hour meters. When the fluo-
rescent tube lights up both will start counting. This is indicated
by the blinking hourglasses.
The hour meter "total h" measures the total time the BiliBed
has been in use. It cannot be reset. By reviewing and separately
recording this display, you have accurate means of determining
when the fluorescent tube should be changed. To ensure
high therapeutical effectiveness, we recommend to change
the fluorescent tube every 1500 hours.
The hour meter "h" shows the therapy time. This meter can
be reset. For resetting to "0": press the START/STOP switch
and hold down for 15 seconds. The BiliBed must be plugged
in. Because this display can be reset, it should not be used
to determine when the bulb should be replaced.
Note
Both displays only light up when the BiliBed is plugged in.
Light Tube Change Log
Lampenwechsel-Kontrolltabelle
Fiche de suivie – Remplacement du tube
Tarjeta de control de cambio de lampara
Actual hours
Date of change
Aktuelle Stunden
Datum
Nb. heures réel
Date
Total horas consumidas
Fecha
Signature
Unterschrift
Signature
Firma