Window Meanings - Varian DRYVAR 150 Manual De Instrucciones

Tabla de contenido
When the MESSAGE has a wrong <CRC>, the
SLAVE will respond with the following string:
<STX>+<ADR>+<NACK>+<ETX>+<CRC>
where <STX>, <ADR>, <ETX> and <CRC> have
the meaning previously discussed, while <NACK> =
0x15.
When the MESSAGE requires an access to an un-
known window, the SLAVE will respond with the
following string:
<STX>+<ADR>+<'2'>+<ETX>+<CRC>
where <STX>, <ADR>, <ETX> and <CRC> have
the meaning previously discussed.
When the MESSAGE is a writing command to a
window, but the data is of a wrong type, the SLAVE
will respond with the following string:
<STX>+<ADR>+<'3'>+<ETX>+<CRC>
where <STX>, <ADR>, <ETX> and <CRC> have
the meaning previously discussed.
When the MESSAGE is a writing command to a
window, but the data has a value out of the admit-
ted range, the SLAVE will respond with the follow-
ing string:
<STX>+<ADR>+<'4'>+<ETX>+<CRC>
where <STX>, <ADR>, <ETX> and <CRC> have
the meaning previously discussed.
When the MESSAGE is a writing command to a
reading only window, or it is a reading command of
a writing only window, the SLAVE will respond with
the following string:
<STX>+<ADR>+<'5'>+<ETX>+<CRC>
where <STX>, <ADR>, <ETX> and <CRC> have
the meaning previously discussed.

Window Meanings

N.
Read/
Data
Write
Type
000
R/W
L
008
R/W
L
100
R/W
L
101
R/W
N
102
R/W
N
103
R/W
N
104
R/W
L
105
R/W
N
108
R/W
N
109
W
L
110
R/W
L
120
R/W
N
121
R/W
N
122
R/W
L
123
Reserved to Varian service
124
33
TECHNICAL INFORMATION
Description
Admitted
Values
Start/Stop
Start = 1
(in remote mode
Stop = 0
the window is a
(default = 1)
read only)
Remote (default)
Remote = 1
or Serial configu-
Serial = 0
ration
(default = 1)
Soft Start
YES = 1
(write only in
NO = 0
Stop condition)
(default = 1 the
first time and 0
the following)
Set Point 1 type
0 = Frequency
1 = Current
2 = Time
(default = 0)
Set Point 1
Hz: 150 to 1050
threshold (ex-
mA: 0 to 500
pressed in Hz,
s: 0 to 4x10
mA or s)
(default = 1000)
9
Set Point 1 de-
0 to 4x10
lay: time between
(default = 0)
the pump start
and the set point
check (seconds)
Set Point 1 out-
0 = output signal
put signal when
high
the measured
1 = output signal
value is greater
low
than the thresh-
(default = 0)
old one
Set point 1 hys-
0 to 100
teresis (in % of
(default = 2)
threshold)
Baud rate
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
Pump life/ cycle
To reset write '1'
time/ cycle num-
ber reset
Interlock type
Impulse = 0
(default = 1)
Continuous = 1
Rotational fre-
150 to 1050
quency setting
(default = 1050)
(Hz)
Maximum rota-
150 to 1050
tional frequency
(default = 1050)
in Hz (active only
in Stop condition)
Set vent valve
On = 1
on/off
Off = 0
(on = closed)
(default = 1)
87-900-920-01(B)
9
Tabla de contenido
loading

Este manual también es adecuado para:

969-9160969-9161969-9162969-9163

Tabla de contenido