Warnings; Maintenance; Cleaning; Sterilization - Carlo De Giorgi 640/00 Manual De Instrucciones De Uso Y Mantenimiento

Electrobisturí
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

ELECTROSURGICAL KNIFE

5.2.4 - Warnings

BEFORE USING THE APPLIANCE CHECK ITS COMPATIBILITY WITH
THE OTHER EQUIPMENT/APPLIANCES IN USE IN THE AREA; THE
ELECTROSURGICAL
KNIFE
GENERATES
ELECTROMAGNETIC
DISTURBANCES DURING NORMAL OPERATION WHICH COULD
INTERFERE WITH THE CORRECT OPERATION OF OTHER DEVICES
IN THE AREA.
ALWAYS SWITCH OFF THE APPLIANCE WHEN YOU FINISH
WORKING
WITH
IT
AND
FOLLOWING
ALL
STOPPAGES,
IRRESPECTIVE OF THE MOTIVE.
IF
YOU
NOTICE
MALFUNCTIONS
STOP
THE
APPLIANCE
IMMEDIATELY AND DISCONNECT IT FROM THE ELECTRICAL
POWER SUPPLY. BEFORE RECONNECTING THE POWER SUPPLY
FIND THE CAUSE OF THE FAULT AND REMEDY IT. IN THE CASE OF
PROBLEMS, CONSULT CARLO DE GIORGI.

6 - MAINTENANCE

6.1 - CLEANING

The appliance does not require particular maintenance with the exception
of regular cleaning of the working area: before cleaning the appliance
always disconnect it from the power supply.
Use non flammable cleaning products and liquids that do not conduct
electricity.

6.2 - STERILIZATION

Pay particular attention to cleaning and sterilizing the handpiece and
electrodes, which can be treated in an autoclave for 13 minutes at 127°C
after bagging.
The cable and neutral electrode must be cleaned with alcohol or sterilizing
products.
IMPORTANT: do not sterilize the handpiece with liquids that conduct
electricity.