5 - OPERATOR INSTRUCTIONS
5.1 - CONTROLS
The appliance is equipped with the following control devices (Fig. 2):
-
illuminated main power ON/OFF switch (1) (green);
-
red AL indicator LED (alarm) (2);
-
plate electrode socket (PLATE) (3);
-
3-position selector (4) with three function LEDs:
CUT
CUT-COAG
COAG
-
output power control potentiometer (5).
The handpiece holder (6) is equipped with a microswitch that deactivates
the appliance when the handpiece is present in the holder.
5.2 - INSTRUCTIONS FOR USE
5.2.1 - Setting into operation
After performing the installation procedures, the appliance can be set into
operation (Fig. 2):
1) fit the required electrode on the handpiece as indicated in § 4.1;
2) select the required function using selector (4);
3) switch on the appliance by means of the main power switch (1);
4) select the required power level by means of potentiometer (5);
5) instruct the patient to grip the neutral electrode;
6) press the button on the handpiece or press the pedal control; this
action will start the audible signal and deliver power to the electrode
attached to the handpiece.
GB - 8