Trouble Shooting Check List - Fiem MG50 Junior Special Manual De Uso

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

TROUBLE SHOOTING CHECK LIST

Before asking for a technical assistance service, please follow this list to try to solve most common problems:
PROBLEM
The machine does not work
Working temperature not sufficient
Egg-turner device not working properly
Forced ventilation not sufficient
Interior lighting not working
Thermometer alcohol
column interrupted
Ref. model MG 100/150 B.F
Not possible to regulate the fan speed
Automatic humidity system does
not work correctly
Semiautomatic humidity system:
Low humidity value
Water loss
Whichever reparation attempt not included on above mentioned form is severely forbidden. Contact always the manufacturer or the authorized tech-
nical assistance in your country. Fiem s.r.l. - Ph : 0039 (0)31 976672 - Fax : 0039 (0)31 899163 - [email protected] - www.fiem.it
CAUSE
Lack of power
Plug disconnected
Master switch "0/I" on position "0"
Safety fuse burnt
Any of the above points
Heating element not correctly working
Thermostat not calibrated
Thermostat faulty or not working
Door open
Cold room under 17/18°C
Egg-turner motor broken
Fan not working
6
Switch on "0" position
Lamp burnt
Accidental drops or jarring of the thermometer
The fan speed regulator does not work
Automatic humidity does not work
properly:
1. broken humidifier
2. plug not inserted
3. hygrostat out of calibration
4. faulty humidity probe
5. no water into humidifier
-Empty external tank
-Natural surface humidification
system does not work correctly
-Water float damaged
-Float does not work correctly
-Float blocked because limestone
SOLUTION
Check the plug
Put the plug into the socket
Position the switch on "I"
Replace the fuse
Contact the assistance service
Contact the manufacturer
See instruction for calibration
Contact the manufacturer
Close the door
Place the incubator into a warmer room
Contact the manufacturer
Contact the manufacturer
Place the switch on "I" position
Replace the lamp as shown
in the instructions
Try to put the thermometer at low
temperature (in a freezer) for few
minutes, and then allow to return
to room temperature.
Otherwise replace the thermometer
Contact the manufacturer
Regarding the point 1.4. contact
the builder;
for the point 2. Insert the plug;
for the point 3. see the paragraph
lcd unit control hygrostat set-up;
for the point 5. check the correct
water system lacing or the
possible lack of water;
-Fill water
-Check the fall down water from
external water to the internal basin
- check the water level setted
from the float
-Check the float
-Clean the float putting a solution of
water 70% + white vinegar 30% inside
the water basin for 2/3 hours.
19
Tabla de contenido
loading

Este manual también es adecuado para:

Mg70/100 familyMg100/150 b.family

Tabla de contenido