Temperature Measurement With Sensor Pt100/Nicr-Ni; Sensor Adjustment (Characteristic Curve Programming) - Festo DIDACTIC Manual De Instrucciones

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

GB
Measuring device with external voltage supply
Temperature monitoring
Press 'E' while carrying out the measurement
to erase the MIN/MAX memory. On completion
of the measurement, press '<' for calling up the
minimum value and '>' for calling up the maxi-
mum value.
/ °Χ
6.5 Temperature measurement
with sensor Pt100 / NiCr-Ni
Connect the sensor to the appropriate test
socket (16 or 17). When using a thermocouple
temperature sensor (SE10), the measuring de-
vice must be operated by battery! An offset
may be input to compensate sensor differences
(zero point offset). Sensing differences can
occur with the Pt100 sensor due to different
contact resistances (example: a contact resist-
ance of approx. 0.5 Ohm results in a zero point
offset of approx. 1.6 Kelvin).
Example: At a known room temperature of 22 ˚C,
the measuring device records 23.6 ˚C with your
Pt100 sensor when uncompensated. An input
of -1.6 is entered to correct your reading.
Key:
ON/OFF
Reading:
---
Key:
E
Reading:
Stor 23.6
Key:
<
Reading:
-1.60
6.6 Sensor adjustment
(Characteristic curve programming)
Example
You want to connect a pressure sensor entail-
ing the following data:
Pressure range: 10...16 bar
Output signal:
resulting in:
Characteristic:
Indicatedvalue2 − Displayvalue1
St =
Measuredvalue2 − Measuredvalue1
Offset:
Off = Indicatedvalue1 − (Measuredvalue ⋅ characteristic)
1 V ⋅ 1,5 bar
= 10 bar −
V
Key:
ON/OFF
Reading:
---
Key:
E
Reading:
Off 0.00
The programmed characteristic can now be
called up:
Key:
S
Reading:
SE1
The sensor must be connected to sockets
SP1+ (10) and SP1- (11).
- 17 -
S
> ...
>
SE1
SE2 ... SE9 OFF0.00
S
E
<
SE9
Off0.00
E
22.0
1...5 V
16bar − 10bar
=
5V − 1V
= 8,5 bar
P
E
>
...
PRO SE11
St 1.00 1.10
>
...
>
E
0.10
8.50
Stor
>
...
>
E
SE2
SE11
XX.XX
E
...
-0,10
bar
= 1,5
V
>
1.50
-----
Tabla de contenido