RECOMMENDED CLEANERS

In general, when used in those concentrations recommended by the manufacturer,
either phenolic type or quaternary (excluding Virex® TB) type disinfectants can
be used. Iodophor type disinfectants are not recommended for use because
staining may result.
Suggested cleaners include:
• Quaternary Cleaners
• Phenolic Cleaners
• Chlorinated Bleach Solution
CAUTION
Do not use Virex® TB to clean the SMRT Charger or SMRT Pak.

CLEANING THE CHARGER

WARNING
• Always wear rubber gloves when cleaning the SMRT Charger to reduce the
risk of injury.
• To avoid electrical hazards, always disconnect the SMRT Charger from the
wall outlet before cleaning.
• Do not spray fluid directly onto the SMRT Charger.
• Do not power wash the SMRT Charger.
• To reduce the risk of electric shock, do not immerse the SMRT Charger in
liquid.
• Do not use solvents, lubricants, or other chemicals to clean the SMRT
Charger unless otherwise directed.
• To reduce the risk of electric shock, do not allow water to collect on top of
the SMRT Charger.
To clean the SMRT Charger:
1.
Disconnect the SMRT Charger from the wall outlet
to avoid electrical hazards during cleaning.
2.
Wipe the surfaces of the SMRT Charger with a
soft cloth dampened with a non-abrasive, hospital
disinfectant (see the Recommended Cleaners list).
3.
Wipe with a cloth moistened with clean water to
remove any cleaning chemicals or residue.
4.
Dry thoroughly before placing the SMRT Charger
back into service.