EN
Instructions for use
- Actuate all movable parts.
- Flush out cannulations and complex mechanisms.
9. Repeat the previous processing steps if visible residue remains present.
10. Perform a final rinse for at least 1 minute using critical water to displace minerals and other impurities
found in utility water. Do not use saline solutions for final rinsing because they may interfere with
disinfection and sterilization.
- Actuate all moving parts
- Pay particular attention to cannulations and blind holes as well as hinges and joints between
mating surfaces.
- Rinse cannulations at least three times with a syringe (volume 1-50ml).
11. Remove excess moisture from the instruments using a clean, absorbent, non-shedding wipe.
12. Allow the instruments to thoroughly dry. Any moisture may affect sterilization and devices may remain
wet after the drying period.
* Manual cleaning was validated using STERIS Prolystica 2X Concentrate Enzymatic Presoak and Cleaner.
PRE-STERILIZATION INSPECTION
• Visually inspect all devices under normal lighting to ensure that cleaning was effective. Pay close
attention to all challenging areas.
- Re-process an instrument that is not clean.
- Replace an instrument that cannot be cleaned.
• Inspect the implants and instruments for surface damage, such as nicks, scratches, and cracks. Replace
any device that is affected.
• Assess the instruments for proper use. Operate all parts and connecting mechanisms. Give careful
attention to drivers, drill bits and reamers, and instruments used for cutting or implant insertion.
Critically assess them for wear, sharpness, straightness, and corrosion. Replace any instrument that
does not perform as intended.
• Inspect all cutting edges under magnification.
- Replace an instrument if a cutting edge is dull, chipped, cracked, rolled, or otherwise deformed.
- Running a cotton cloth over the edge may help detect chipping and cracking.
• Verify the legibility of all markings and reference scales. Replace any device that is unreadable.
• Repair, replace, and/or repeat the cleaning of instruments as needed to ensure proper operation before
proceeding with sterilization.
• Lubrication ("instrument milk") may increase the useful life of surgical instruments. Do not use silicone-
based lubricants, oil, or grease, as these will interfere with steam sterilization. Only use a water-based
lubricant intended for use on surgical instruments and with steam sterilization. Use the lubricant as
directed by the manufacturer. Use critical water if dilution is required.
• Fully replenish the system trays and caddies.
STERILIZATION
• Perform sterilization using a dynamic-air-removal (prevacuum) autoclave.
- Gravity displacement sterilization is not recommended.
- Immediate use (flash) sterilization is not recommended.
• Ensure the sterilizer's maximum load limit is not exceeded when sterilizing multiple sets or devices.
• Do not stack containers as this might prevent the penetration of steam and inhibit drying.
• Refer to the sterilizer manufacturer's instructions and ensure proper installation, calibration, use, and
ongoing maintenance.
English (non-U.S.) – EN / PAGE 47