Section 9 Depth sensors
9.1 Specifications
Specifications are subject to change without notice.
Specification
Range
Accuracy
Resolution
Parameters
Maximum depth
Operating temperature
9.2 Product overview
The depth sensor is in the sonde and cannot be seen.
9.3 Install a seal screw
Do not deploy a sonde that has a 25 m depth sensor to a depth of more than 35 m (164 ft) unless a seal screw is
installed in the sensor port or damage to the sensor will occur.
Install a seal screw in the sensor port of a sonde with a 25 m depth sensor before the sonde is
deployed to more than 35 m (164 ft). Refer to Figure
for ordering information.
Remove the seal screw to enable the depth sensor to operate.
Figure 10 Install a seal screw
9.4 Accessories
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
6
To prevent damage to the depth sensor, install a seal screw in the sensor port before
deployment of more than 35 m (164 ft).
25 m
0 to 25 m (0 to 82 ft)
± 0.05 m
0.01 m
Depth (meters or feet), psia, density, specific gravity (calculated)
6
35 m (164 ft)
-5 to 50 °C (23 to 122 °F), non-freezing
N O T I C E
W A R N I N G
100 m
0 to 100 m (0 to 328 ft)
± 0.05 m
0.01 m
225 m (738 ft)
Figure
10. Refer to
Accessories
200 m
0 to 200 m (0 to 656 ft)
± 0.1 m
0.1 m
225 m (738 ft)
on page 23
English 23