"m.v." means a standard measured value.
True RMS measurement for AC current
Range
200.0 A
All AC current ranges are specified from 5% to 100% of range
Frequency range: 50 Hz...60 Hz
True RMS voltage measurement
Range
6.000 V
60.00 V
600.0 V
1000 V
All AC voltage ranges are specified from 5% to 100% of range
Input impedance: 10 MΩ
Frequency range: 50 Hz...1000 Hz (sine wave), 50/60 Hz (all
waves)
Overload protection: 1000 V DC/AC RMS
DC voltage measurement
Range
6.000 V
60.00 V
600.0 V
1000 V
Input impedance: 10 MΩ
Overload protection: 1000 V DC/AC RMS
11 Technical data
Resolution
0.1 A
Resolution
0.001 V
0.01 V
0.1 V
1 V
Resolution
0.001 V
0.01 V
0.1 V
1 V
CMP-200F – USER MANUAL
Accuracy
(3% m.v. + 5 digits)
Accuracy
(1.2% m.v. + 5 digits)
(1.2% m.v. + 2 digits)
(1.5% m.v. + 2 digits)
Accuracy
(0.9% m.v. + 5 digits)
(1.0% m.v. + 2 digits)
(1.2% m.v. + 2 digits)