WIKA GFU08 Manual De Instrucciones página 22

Carro de servicio para gas sf6
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

6. Commissioning, operation
Protect the gas cylinder valve by an appropriate valve protection.
To test the leak tightness, a model GIR-10 leak detector or a comparable leak
detector must be used.
Close the outlet valve at the pressure reducer.
EN
After opening the gas cylinder valve, the initial pressure is available at the pressure
reducer.
6.5 Filling and refilling of SF
6.5.1 Connecting the filling hose
Establish a gas-tight connection between the SF
compartment to be filled using the filling hose included.
6.5.2 Adjusting the pressure reducer
Use the pressure setting handle to set the pressure reducer to the desired output
pressure. For this, read the pressure on the output pressure gauge.
6.5.3 Weighing the quantity of the filled SF
with scales)
CAUTION!
Place the gas cylinder without the shunt forces of the fastening elements
(ratchet strap or chain) on the scale plate.
Secure the gas cylinder against falling over!
Place the SF
gas filling cart level on solid ground in order to make a precise weight
6
measurement.
Correct position of the gas cylinder on the scale plate:
22
gas
6
gas filling cart and the gas
6
gas (models GFU08-W and GFU08-C
6
WIKA operating instructions SF
gas filling cart, model GFU08
6
Tabla de contenido
loading

Tabla de contenido