CALIBRATION GRAVITY
The Scale displays the calibrated gravity, e.g. G9.794 for g = 9,794.
Press
, to enter the Gravity setting. G---- shows.
Press
to modify the digit. Press and hold
Press and hold
to input the decimal point.
MAX. CAPACITY
Do NOT attempt to set Max. Cap. greater than scale's actual capacity. Overloading
causes severe harm to scale, and is very dangerous.
Press
, to enter the Max. Cap. setting. 00000 shows.
Press
, to modify the digit. Press and hold
capacity value. Press and hold
to input the decimal point.
ZERO DETECTION
Press
, to enter the zero detection. LOAD0 shows.
Keep the scale without load and press
EN
to go to the next step.
LOAD 1 DETECTION
The scale displays „LOAD1".
Load a standard weight, and press
. „00000" shows
Press
, to modify the digit. Press and hold
to input the decimal point.
Keep the load stable and press
to display the weight code 23456.
Wait until the weight code is stable. Press
Once the detection finished, the scale automatically enters Load2 Detection.
LOAD 2 DETECTION
The scale displays LOAD2 .
REV. V 2016
to move the digit. Input the Gravity value.
to move the digit. Input the max.
, to display the weight code 12345. After the code is stable press
to move the digit. Input weight value. Press and hold
to start the weight detection.
24
Load a standard weight, and press
. „00000" shows.
Press
, to modify the digit. Press and hold
and hold
to input the decimal point.
Keep the load stable and press
, to display the weight code 34567.
Wait until the weight code is stable. Press
to start the weight detection.
Once the detection finished, the scale automatically enters Load3 Detection.
LOAD 3 DETECTION
The scale displays LOAD3.
Load a standard weight, and press
. „00000" shows.
Press
, to modify the digit. Press and hold
Press and hold
to input the decimal point.
Keep the load stable and press
to display the weight code 45678.
Wait until the weight code is stable. Press
to start the weight detection.
Once the detection finished, the scale automatically saves and exits the Calibration mode.
Problem
Possible Cause
discharged / defective battery
not power-on after
is pressed
defective
key
defective power cable
defective mainboard
display flashes
discharged battery
scale is disturbed
no action taken after key
pressed
defective key
weight reading not stable
load in motion
weak Anti-Motion
damped loadcell or mainboard
defective mainboard
REV. V 2016
to move the digit. Input weight value. Sie das GePress
to move the digit. Input weight value. Sie das Gewicht
Suggested Solution
check batteries and charge / exchange
the batteries
press harder and keep pressing 2sec
open front panel, check power cable
contact service representative
charge / exchange the batteries
Turn off and then turn the scale on
again
contact service representative
keep load stable
change Anti-Motion level
dry loadcell or mainboard
contact service representative
25
EN