Steinberg Systems SBS-KW-1000/10G Manual De Instrucciones página 11

Ocultar thumbs Ver también para SBS-KW-1000/10G:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 33
DISPLAY BRIGHTNESS
Press
, to enter Display Brightness. br shows.
Press
, to change brightness value.
Display Brightness can be set to:1(dim), 2(normal), 3(bright).
Brightness saves battery power.
IDLE MODE TIMING
Press
, to enter Idle Mode Timing. Idl shows.
Press
, to change timing value.
Idle Mode Timing can be set to: 0(disabled), 5(5sec), 10(10sec), 15(15sec), 30(30sec), 60 (60sec). It is 30sec
by default.
To maximize battery life, the scale automatically enters Idle Mode, when there's no action or the load is
stable. In Idle Mode, the scale works in low-power consumption status. Any key pressing or motion in load
wakes up the scale from Idle Mode.
Press
, to exit User Setup.
SCALE CONFIGURATION
EN
ATTENTION: The parameters in Scale Configuration are closely related to the scale's
metrology performance. It is NOT recommended to change any parameters unless
you absolutely need to. In case you´re not sure, please contact the customer service to
check if it is necessary to modify these parameters.
SCALE CONFIGURATION
ENTERING THE MENU
Simultaneously press
,
, and
The Scale displays P0000. . Input Scale Configuration password P0258 . To do so, press
digits. To move the digits, press and hold
Press
to confirm the input and enter the Scale Configuration.
Scale displays SCALE.
DISPLAY RESOLUTION
The display resolution can be set to: The display resolution can be set to: 0.001, 0.002, 0.005, 0.01, 0.02, 0.05,
0.1, 0.2, 0.5, 1, 2, 5, 10, 20, or 50.
Press
to enter the Display Resolution. E--- shows.
Press
to change the resolution value.
REV. V 2016
to enter Password mode.
.
20
AUTO-ZERO RANGE
The Auto-Zero range can be set to: 0% (disabled), 2 (±2%FS), 3 (± 3%FS), 4 (±4%FS), 10 (±10%FS), 20
(±20%FS), 50 (±50%FS), 100 (±100%FS). It is set to ±20%FS by default.
Press
to enter the Manual-Zero range setting. The scale displays the Manual-Zero range,
indicating 4% F.S. Zero is only allowed when the weight is within the Manual-Zero range.
Press
, to change the value.
MANUAL ZERO RANGE
The manual zero range can be set to: 0% (disabled), 2 (±2%FS), 3 (±3%FS), 4 (±4%FS), 10 (±10%FS), 20
(±20%FS), 50 (±50%FS), 100 (±100%FS).
Press
to enter the Manual-Zero range setting.
The scale displays the Manual-Zero range,
within the Manual-Zero range.
Press
ZERO TRACKING RANGE
The Zero-Tracking range can be set to: 0 (disabled), 0.5 (±0.5e), 1.0 (±1.0e), 1.5 (±1.5e), 2.0 (±2.0e), 2.5
(±2.5e), 3.0 (±3.0e), 3.5 (±3.5e), 4.0 (±4.0e), 4.5 (±4.5e), 5.0 (± 5.0e). Enabling the Zero-Tracking will enhance
the scale tempe¬rature and drift the performance!
Press
,to enter the Zero tracking range setting. The scale displays the Zero-Tracking range
The scale displays active value, e.g. 0.5 for the 0.5e value.
to modify the
Press
, to change the value.
REV. V 2016
indicating 4% F.S. Zero is only allowed when the weight is
to change the value.
.
21
EN
Tabla de contenido
loading

Este manual también es adecuado para:

10030128

Tabla de contenido