Safety Features
The following information will acquaint you with
safety features of the VisionAire Oxygen Concen-
trator. Make sure you read and understand all the
information contained in this manual before you
operate your unit. Should you have any questions,
your Equipment Provider will be happy to answer
them for you.
Device warning label and alarm display.
• Compressor Motor: A pressure relief valve is
fitted to the compressor outlet and is calibrated to
280 kPa (40 psig). Thermal safety is ensured by
a thermostat situated in the stator winding of the
compressor (135°C / 275 °F).
• General Malfunction: If any of the conditions
listed below occurs, the general malfunction light
(
) will illuminate and an audible intermittent
alarm will activate.
This includes:
• Obstruction to the flow of oxygen such as a
pinch or kink in the delivery cannula, trig-
gered by high product tank pressure
• High device product tank pressure condition
of greater than 33psig (±1)
• Low device product tank pressure condition
of less than 5psig (±1)
• High device temperature of greater than
135°C (275 °F), triggered by low product
tank pressure if the thermal switch located
within the compressor trips (shutting down
the compressor)
VisionAire Family
• Oxygen Monitor: The oxygen monitor detects
any drop in concentration below 82%. If this oc-
curs the low oxygen concentration warning light
(
) will illuminate. If the low O2 condition
persists, an audible intermittent alarm will also
activate in addition to the alarm light.
• Power Failure: In the event the unit is operating
and a loss of power occurs, the power warning
light (
) will illuminate and an audible
intermittent alarm will activate. See label image.
• Product Filter: ≥ 10 μm filter
PN MN238-C4 C | User Manual