Defective Platform; Unfortunately, The 3D L.a.s.a.r. Posture App Has Stopped - Ottobock 3D L.A.S.A.R. Posture Instrucciones De Uso

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 257
Use

5.4.9 Defective platform

Warning
Defective platform
Measuring accuracy too low.
The platform is defective.
Please contact your local customer service.
Realign the measurement plate.
0.0 Nm
0.0 %
Defective platform
The platform is defective.
Please contact your local customer service.

5.4.10 Unfortunately, the 3D L.A.S.A.R. Posture app has stopped.

Apps
File Manager
Unfortunately , the 3D L.A.S.A.R. Posture app has stopped.
Unfortunately, the 3D L.A.S.A.R. Posture app has stopped.
102
Live view
Calibration failed. Tap to recalibrate...
Cancel
O.K.
Recalibrate
0.0 Nm
0.0 %
1
2
O.K.
Widgets
Settings
OK
3D L.A.S.A.R.
Posture
1
2
Error message text:
: "Defective platform"
: "The platform is defective."
: "Please contact your local customer service."
: "OK"
Possible causes:
Broken cable
Torn cable
Defective circuit board
Measures:
► Have the defective force measurement plate
repaired by Ottobock Service.
► 1x tap on "OK".
Error message text:
: "Unfortunately, the 3D L.A.S.A.R. Posture app has
stopped."
: "OK"
Possible causes:
Unexpected app error
Measures:
► 1x tap on "OK" and restart the app.
► Contact Ottobock Service if this problem recurs
often.
OK
743L500 3D L.A.S.A.R. Posture
Tabla de contenido
loading

Este manual también es adecuado para:

743l500743l500 3d l.a.s.a.r. posture

Tabla de contenido