Gebruik
5.4.9 Platform defect
Warning
Defective platform
Measuring accuracy too low.
The platform is defective.
Please contact your local customer service.
Realign the measurement plate.
0.0 Nm
0.0 %
Defective platform
The platform is defective.
Please contact your local customer service.
5.4.10 De 3D L.A.S.A.R. Posture app is helaas gestopt.
Apps
File Manager
Unfortunately , the 3D L.A.S.A.R. Posture app has stopped.
Unfortunately, the 3D L.A.S.A.R. Posture app has stopped.
356
Live view
Calibration failed. Tap to recalibrate...
Cancel
O.K.
Recalibrate
0.0 Nm
0.0 %
1
2
O.K.
Widgets
Settings
OK
3D L.A.S.A.R.
Posture
1
2
Tekst van de foutmelding:
: "Platform defect"
: "Het platform is defect."
: "Neem contact op met uw lokale servicedienst."
: "OK"
Mogelijke oorzaken:
•
Kabel gebroken
•
Kabel afgescheurd
•
Printplaat defect
Maatregelen:
► Laat de defecte krachtmeetplaat door Ottobock
Service repareren.
► "OK" 1x aantikken.
Tekst van de foutmelding:
: "De 3D L.A.S.A.R. Posture app is helaas gestopt."
: "OK"
Mogelijke oorzaken:
•
Onverwachte fout in de app
Maatregelen:
► "OK" 1x aantikken en de app opnieuw opstarten.
► Neem contact op met Ottobock als dit vaker
optreedt.
OK
743L500 3D L.A.S.A.R. Posture