1.
Screen
2.
Spirit level
3.
Lamp of control
4.
ON / OFF
5.
Platform
6.
Doorknob
7.
Right glass
8.
Insulation
9.
Left glass
10. Top Glass
11. Foot
10

3. CONFIGURATION

MODE: Confirmation button
ZERO: Choice button
a) Options of the backlight: AUTO BLK, ON, OFF
b) Options of print: 0CoM, 1COM, 2COM
c) Baud-Rate: 9600, 2400, 4800
4. OPERATION
1.
Preheat Turn on the scale. The screen displays „0000000" and goes to „9999999".
The internal number of the model is displayed, then „0.00". Let the scale preheat
for at least 15min to reach the most accurate results.
2.
Calibrate see the corresponding chapter.
3.
Weighing If the scale is overloaded, it displays „HHHHHH" and issues a warning
signal. Remove the weight of the platform immediately.
4.
Tare if „0" flashes during the use or if the weight of the glass appears on the
screen, press TARE to reset the scale to zero.
5.
Counting function Put a unit on the platform and press the MODE button to
switch to the counting function. You can choose between two methods of coun-
ting.
1. „PCS" = count
2. „PCT" = percentage
The scale now displays the weight of the unit. Press TARE and ZERO to increase
to lower the percentage value. The value indicated flashes and can be adjusted
with the TARE button. (100 / 200 / 400 / 500/600/700/800/900/1000/10/20/30/40/
50/60 / 70/80/90). The number chosen is the number of units placed on the scale.
Confirm with PCS. Then place the pieces to count on the scale, which displays
the corresponding number.
5. CALIBRATE
To calibrate, let preheat the scale as shown above. When zero is flashing, press the
TARE button: the scale displays „0".
Once the figure is stable, press the ZERO button for 5 sec. until the display shows
„CAL". The scale displays „EMPTY"
Wait until the numbers have stabilized: the balance is then zeroed.