► Follow the safety instructions to avoid injuries and damage to the
product.
► Instruct the user in the proper and safe use of the product.
► Please keep this document in a safe place.
1.1 Construction and Function
The 5R9 socket attachment block is bonded and laminated to the
foamed distal end of the prosthetic socket. The 4R60 socket adapter
serves to connect with the distal prosthetic components and is
fastened to the 5R9 socket attachment block.
1.2 Combination possibilities
This prosthetic component is compatible with Ottobock's system of
modular connectors for children. Functionality with components of
other manufacturers that have compatible modular connectors has not
been tested.

2 Intended use

2.1 Indications for use
The product is to be used solely for lower limb prosthetic fittings.
2.2 Area of application
Approved for a body weight up to max. 45 kg.
Approved for a body height up to max. 145 cm.
Approved for a foot size up to max. 21 cm.
2.3 Environmental conditions
Storage and transport
Temperature range –20 °C to +60 °C (–4 °F to +140 °F), relative
humidity 20 % to 90 %, no mechanical vibrations or impacts
Allowable environmental conditions
Temperature range: –10 °C to +45 °C
Moisture: relative humidity: 20% to 90%, non-condensing
8
Unacceptable environmental conditions
Chemicals/liquids: fresh water, salt water, perspiration, urine,
acids, soapsuds, chlorine water
Solids: dust, sand, highly hygroscopic particles (e. g. talcum)
2.4 Service life
In general, all modular adapters are tested by the manufacturer for
3 million load cycles. Depending on the user's activity level, this cor­
responds to a maximum service life of 5 years.
3 Safety
3.1 Explanation of warning symbols
Warning regarding possible risks of accident or
CAUTION
injury.
3.2 General safety instructions
CAUTION
Excessive strain on the product
Risk of injury due to breakage of load-bearing components
► Use the product according to the specified area of application
(see Page 8).
CAUTION
Unallowable combination of prosthetic components
Risk of injury due to breakage or deformation of the product
► Only combine the product with prosthetic components that are
approved for that purpose.
► Based on the instructions for use of the prosthetic components,
verify that they may be combined with each other.
Tabla de contenido
loading

Este manual también es adecuado para:

4r60

Tabla de contenido