If, during this battery charge process the equipment is connected to start measurements, the charge process will be tempo-
rarily interrupted. If the red LED ? is lit, it will turn off and all the charge will be used to feed the equipment. If the green
LED ? is on, it will remain lit to show the battery is charged (though it ceases to charge it). Upon completing measure-
ments or when the apparatus is disconnected, the charge process will automatically resume.
Note: The battery looses part of its charge when stored. Therefore, the battery should be recharged before using the megger
for the first time or after a given period without being used.
Technical specifications
Test voltages (T.V.)
Maximum resistance
reading
Short circuit current
Test voltages accuracy
Megohmmeter basic
accuracy
Advanced features
Built-in printer
Serial data output
Built-in chronometer
Environmental protection
Safety class
E.M.C
Electrostatic immunity
Electro magnetic irradiation
immunity
: 500, 1000, 2500, 5000 V directly, one button selectable.
500 V to 5 kV in 500 V steps.
DC, negative.
: 5TΩ @ 5 kV
: 1,5 ± 0,5mA
: ± 3% of nominal test voltages on R ≥ 10GΩ
: ± 5% of reading ± 3 digits
(1MΩ to 500GΩ at any test voltage)
: • Automated Polarization Index computing
• Automated Dielectric Absorption Ratio computing
• Pass fail test
: Prints elapsed time, actual voltage and resistance measured.
: RS-232 @ 4800 bps. Suitable for data collection in an external serial printer, computer or
data-logger.
: Shows elapsed time in mm:ss format. Count starts automatically for each measurement.
: IP54 with closed lid.
: Meets the requirements of IEC 61010-1/1990, IEC 61010 1/1992 amendment 2
: In accordance with IEC 61326-1
: In accordance with IEC 1000-4-2
: In accordance with IEC 61000-4-3
34