Circutor MD-5060e Guia Del Usuario página 27

Megóhmetro digital
Ocultar thumbs Ver también para MD-5060e:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

9.
Press key * and select the Voltage Test with any of the 2 to 5 keys. If you want other Voltage Test between
500V and 5kV, press 6 or 7 key.
10.
If the key * is pressed again the voltage test selected will be hold, but it is not necessary.
11.
Press the START key 8. Shall the "HIGH VOLTAGE" light indicator C will be on, indicating that the megohmmeter
internal generator is applying the selected voltage to the element being tested. The = LED will turn off.
12.
For a few seconds, the self-scale smart system will search for the most convenient range for the value being meas-
ured. Meanwhile, the display will show the message "WAIT..." Then, the display will show during a few seconds , the
number of test and then the measurement result using both digital (measurement value and unit) and analog displays.
13.
If, for example, the measure value was 601 MΩ with 1.000 V voltage test, it will be read.
14.
If the measured value exceeds 5000 GΩ @ 5 kV, it will be read the following message:
29

Capítulos

Tabla de contenido
loading