Figure 9 Floating point definition
1 Sign bit
2 Exponent
3 Mantissa
Word wise swapping
In word wise swapping, the third and fourth bytes are interchanged in
order with the first and second bytes. This results in a byte order of
3 4 1 2. The byte order complies to the IEEE Big Endian floating point
definition.
Troubleshooting
Error and status indicators
Error and status words follow the same standard definition for all sc
probes and controllers.
Table 7
lists bit position and error messages.
and status messages.
A bit value of zero shows the error or status condition that is not true.
4 Exponent
5 Mantissa
Table 8
lists bit position
A bit value of 1 shows the error or status condition that is true. For
example, if Bit 0 has the value of 1, an error has occurred during the last
calibration.
Table 7 Error messages
Bit Message
0
Measurement calibration
error
1
Electronic adjustment error
2
Cleaning error
3
Measuring module error
4
System re-initialization
error
5
Hardware error
6
Internal communication
error
7
Humidity error
8
Temperature error
9
—
10
Sample warning
11
Questionable calibration
warning
12
Questionable
measurement warning
13
Safety warning
Indication
An error has occurred during the last
calibration
An error has occurred during the last
electronic calibration
The last cleaning cycle failed
A failure has been detected in the
Measurement Module
Some settings are inconsistent and have
been reset to factory defaults
A general hardware error has been detected
A communication failure within the device has
been detected
Excessive humidity has been detected within
the device
Temperature within the device exceeds a
specified limit
—
Some action is required with the sample
system
The last calibration may not be accurate
One or more of the device measurements are
out of range or are of questionable accuracy
A condition has been detected which may
result in a safety hazard
English 13