• The cuff is fastened too tightly or loosely: E3;
• Errors occur during the measurement: E4;
• The pump pressure is higher than 300 mmHg: E5;
• The batteries are almost empty
In such cases, repeat the measurement. Ensure that you do not
move or speak. If necessary, reinsert or replace the batteries.
8. Cleaning and Care
• Clean your blood pressure monitor carefully using a slightly
damp cloth only.
• Do not use any detergents or solvents.
• Never hold the instrument under water as otherwise liquid
can penetrate and damage the instrument.
• Never place any heavy objects on the instrument.
9. Technical details
Model no.
BC 32
BC32/1
Type
Measurement
Oscillometric, non-invasive blood
method
pressure measurement on the wrist
Measurement range
Cuff pressure 0 – 300 mmHg,
systolic 50 – 250 mmHg,
diastolic 30 – 200 mmHg,
Pulse 40 –180 beats/minute
Display accuracy
Systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
E6.
18
Measurement
Max. permissible standard deviation
inaccuracy
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Memory
2 x 60 memory spaces
Dimensions
L 72 mm x W 69 mm x H 29.2 mm
Weight
Approx. 101 g (without batteries)
Cuff size
135 to 195 mm
Permissible operating
+5 °C to +40 °C, 15-93 % relative air
conditions
humidity (non-condensing)
Permissible storage
-25 °C to +70 °C, 10-93 % relative air
conditions
humidity, 700 –1060 hPa ambient pres-
sure
Power supply
2 x 1.5 V
Battery life
For approx. 250 measurements,
depending on the blood pressure level
and/or pump pressure
Accessories
Instruction for use, 2 x 1.5 V AAA batter-
ies, storage box
Classification
Internal supply, IPX0, no AP or APG,
continuous operation, type BF applied
part
Technical information is subject to change without notification
to allow for updates.
AAA batteries