Ottobock Myo Plus Instrucciones De Uso página 29

Ocultar thumbs Ver también para Myo Plus:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 60
Table 4 - casing
Phenomenon
Electrostatic dis­
charge
High frequency elec­
tromagnetic fields
High frequency elec­
tromagnetic fields in
the immediate vicinity
of wireless communic­
ation devices
Magnetic fields with
rated power frequen­
cies
Table 5 – alternating current gate for the supply input
Phenomenon
Electrical fast transi­
ents/bursts
Surges
Line against line
Conducted interfer­
ence induced by high-
frequency fields
Voltage drops
Voltage interruptions
EMC basic standard
or
test procedure
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-8
EMC basic standard
or
test procedure
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
IEC 61000-4-11
Interference immunity test level
Professional health­
care facility
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air,
3 V/m
80 MHz to 2.7 GHz
80 % AM at 1 kHz
See Table 9
30 A/m
50 Hz or 60 Hz
Interference immunity test level
Professional health­
care facility
± 2 kV
100 kHz repetition rate
± 0.5 kV, ± 1 kV
3 V
0.15 MHz to 80 MHz
6 V in ISM frequency
bands between
0.15 MHz and 80 MHz
80% AM at 1 kHz
0 % U
; 1/2 period
T
at 0, 45, 90, 135, 180, 225, 270 and 315
degrees
0 % U
; 1 period
T
and
70 % U
; 25/30 periods
T
Single phase: at 0 degrees
0 % U
; 250/300 periods
T
Environments in
areas of home
healthcare
10  V/m
80 MHz to 2.7 GHz
80 % AM at 1 kHz
Environments in
areas of home
healthcare
3 V/
0.15 MHz to 80 MHz
6 V in ISM and ama­
teur radio frequency
bands between
0.15 MHz and 80 MHz
80% AM at 1 kHz
29