axillary mode:
rectal mode:
direct mode
(all measuring):
typical duration before switching to the direct mode:
at an ambient temperature
over 35 °c (95 °f):
if no measuring point
is recognised:
if the temperature
does not stabilise:
Water bath accuracy (between 35.5 °c and 42.0 °c):
all measuring points:
thermometer probe:
insert:
materials:
probe shaft:
probe cable:
probe tip:
all materials of the predictiVe thermometer are latex-free.
ri-vital
®
technIcaL Data
68
8 - 12 seconds
10 - 14 seconds
60 - 120 seconds
Immediate
60 seconds (after removing the probe from the probe slot)
70 seconds (after placement at the measuring point)
±0,1 °c (±0,2 °F)
replaceable
removable, replaceable, washable, Pc+aBs
Probe handle, Flame retardant polycarbonate/polyester
mix
highly fire-resistant polyester
Polyurethane sheath with tPe overmould
aluminium
Body temperature measurement with the optional ear thermometer:
the device can optionally be equipped with a holder for the
in place of the predictive thermometer.
this holder allows convenient storage of the thermometer and the protective sleeves
required for the measuring head.
For more information about using the device and important instructions regarding
measuring precision, please refer to the operating manual for the
12.3 Blood pressure (niBp)
Disinfectable bladderless cuffs: all cuff sizes from the riester portfolio
measurement range:
systole 25 - 280 mmhg
Diastole 10 - 200 mmhg
MaP 15 - 260 mmhg
measurement range
for neonates:
systole 20 - 155 mmhg
Diastole 5 - 110 mmhg
MaP 10 - 130 mmhg
measurement precision:
Mean value difference < +/-5mmhg
standard deviation < 8mmhg
< 6 mmhg /minute.
leakage:
overpressure limit:
300 mmhg mode „adult"
150 mmhg mode „neo" (neonates)
Overpressure limit (2nd Level): 330 mmhg mode „adult"
165 mmhg mode „neo"(neonates)
time per measurement
approx. 25s on average „adult" mode max. 170 s
„neo" mode max. 80s
ri-vital
®
technIcaL Data
ri-thermo
nthermometer
®
ri-thermo
n
®
69
Tabla de contenido
loading