Standards Complied With - Riester ri-vital Instrucciones De Uso

Ocultar thumbs Ver también para ri-vital:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

electromagnetic stability (immunity), portable rf devices
the distance between mobile communication devices and all parts of the device
including all cables should not fall below the distance calculated with the
corresponding equation for the frequency of the communication device.
test
iec 60601-1-2
emitted energy
3V/m
3V/m
Grid-bound
energy
3 VrMs
the field strengths of stationary senders, such as base stations for cellular phones, of land mobile radio, amateur
radio, aM and FM radio transmissions and tV broadcasts cannot be predicted.
a site survey should be conducted for assessing the electromagnetic environment of stationary rF transmitter.
If the sender's measured field strength at the device's usage site exceeds the above-mentioned values, then the
device's function should be monitored.
If monitoring shows that the device is not functioning properly, corresponding measures must be taken. For
example, the device must be realigned or placed at a different location.
recommended distances
the following minimum distances are recommended between the device and portable hF communication
equipment (e.g. mobile phones):
hf frequecy
26Mhz - 80Mhz
equation
d=1.2*square root (P)
nominal power (W)
distance (m)
0.1
0.40
1
1.20
10
3.80
50
8.50
P is the nominal power of the transmitter in watts, d is the distance in metres
example:
at a transmission frequency of 700 Mhz and power of 10 watts, a distance of d = 1.2*root(10) = 3.79 m is
recommended.
ri-vital
®
eLectrOMaGnetIc staBILItY
76
class
distance
3V/m
Distance =1.2*square root (power) for 80.. 800 Mhz
3V/m
abstand = 2.3*square root (power) for 800 Mhz to
2.5 Ghz Iec 61000-4-3
3 VrMs
square root (power) for 150khz to 80Mhz
Iec 61000-4-6
80Mhz - 800 Mhz
800Mhz - 2.5 Ghz
d=1.2*square root (P)
d=2.3*
square root (P)
distance (m)
distance (m)
0.40
0.80
1.20
1.20
3.80
7.30
8.50
16.30

standards complied with:

harmonic emissions
Iec 61000-3-2
Flicker emissions
Iec 61000-3-3
Immunity from table
Iec 61000-4-2, level 3,
electrostatic devices
discharge
Immunity form electromagnetic
Iec 61000-4-3 at 3V/m
fields caused by
high-frequancy radiation
Immunity from sudden voltages
Iec 61000-4-5, level 3
peaks/surges
current surge immunity
Iec 61000-4-5, level 3
electromagnetic
Iec 61000-4-6 at 3V/m
compatibility caused
by conduction
Power frequency
Iec 61000-4-8 at 3 V/m
magnetic fields
Operation durin mains
Iec 61000-4-11, table 7
voltage flucations
Warning:
using accessories other than those specified can lead to increased emissions and/or decreased stability of the
device.
nominal power (w)
Distance (m)
16. standards complied with:
Basic norm:
en 60601-1:1990 + a1:1993 + a2:1995 + a13:1996
emV:
en 60601-1-2:2001 + a1:2006
spo2:
IsO9919:2009
niBp:
en 1060-3:1997+a2:2009
en 1060-4:2004
en 1060-1:1995 + a2:2009
thermometer:
en 12470-3:2000+a1:2009
ri-vital
®
stanDarDs cOMPLIeD wIth
77
Tabla de contenido
loading

Tabla de contenido