WIKA DIH10 Manual De Instrucciones página 14

Cabezal con indicador digital integrado
Ocultar thumbs Ver también para DIH10:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

5. Commissioning, operation
5.1 Operating a current loop
Power
EN
supply
unit
DC 24 V
The display is only ever operated in a 4 ... 20 mA current loop. A direct
connection to a voltage supply (e.g. DC 24 V) will lead to the display
being damaged and will invalidate the warranty.
For a functional check, the display must be powered from a mA source
(e.g. mA transmitter or calibrator).
In normal operation, the display is connected in series with a 4 ... 20 mA
transmitter or connected to the 4 ... 20 mA analogue output of an
instrument.
14
internal
bridge
DIH10
display
Case
WIKA operating instructions model DIH10
Tabla de contenido
loading

Tabla de contenido