Troubleshooting - Heidolph LABOROTA 20 Instrucciones De Empleo

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

E
Master switch won't light when turning ON
check power cord
master switch broken (call service)
Display: Act Vacuum "0000"
vacuum sensor broken or not connected
Display: Act Vacuum won't change during evacuation
vacuum valve connected with vacuum pressure sensor terminal
Display: Act Vapor "----"
temperature sensor not connected or broken
Display: Act T auto "----"
T auto temperature sensor not connected or broken
Display: Act Chiller "----"
Cooling water temperature sensor not connected or broken
Vacuum valve inoperative
vacuum pressure sensor not connected with valve terminal, or valve broken
Display: Act Vacuum (aerated) differs from atmospheric
altitude correction factor not integrated.
Calibrate vacuum pressure sensor.
Poor vacuum pressure
inspect gaskets and fittings
check ground fittings, coat with vacuum grease, as required
PTFE seal installed the wrong way. Sealing lip needs to point towards drive head
check line to vacuum pump
Heating bath lift won't travel up
float switch not connected or broken
also refer to safety functions section
lift mechanism broken
upper limit switch broken
Heating bath lift won't travel down
lift mechanism broken
lower limit switch broken
No rotation
also refer to safety functions section
motor broken
Mode won't change from P const to T auto.
T auto temperature sensor not connected or broken
Aerating valve won't work. Aerating valve broken or not connected.
Distillation automatically discontinued (by accident)
EndP value incorrect; Set Time incorrect, cut-out sensor responded.
Also refer to safety functions (page 95)
Whenever facing a problem that cannot be fixed by the above information, call your authorized
Heidolph Instruments dealer right away.

TROUBLESHOOTING

110