Disposal; Troubleshooting; Specifications - Heidolph REAX 2 Instrucciones De Empleo

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

E
Caution
Do not jolt or shake the item during transport.
For disposal, please comply with your local or national regulations.
Split by metal, plastic, etc.
Packing material to be treated as described above (material split).
Work on electric and electronic components is limited to qualified personnel.
Master switch on shaker / mixer won't light
1. check power cord
2. check circuit breakers
Item won't shake (master switch lighted))
1. Thermal motor circuit breaker triggered by motor overload
Remedy:
Wait about 20 minutes, decrease load applied on shaker plate.
2. Mechanical parts broken (humming motor noise) or electronic failure (no motor noise).
REAX 2
power supply
input power
shaker frequency
mixing movement
drive
adapter
Load-bearing capacity
weight
protection class
ambient temperature
space required
All manuals and user guides at all-guides.com

DISPOSAL

TROUBLESHOOTING

SPECIFICATIONS

230/240 V 50/60 Hz / 115 V 50/60 Hz (optional)
27 W
20 – 100 1/min
overhead rotation
split-pole motor, features electronic speed control;
motor overheating protection with automatic reset
Interchangeable, see accessories
max. 1 kg, limitations for off-center load
4,8 kg
IP 21
0°C to 50°C at 80% rel. humidity
Approved for installation in gassing and conditioning
cabinets (make reference to temperature limits)
460 mm x 175 mm