Error Codes List; Service Information; Technical Data; Ambient Conditions - OHAUS CW-11 Manual De Instrucciones

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

CW-11 Indicator

5.2Error Codes List

The following list describes the various error codes and which can appear on the display and the suggested
remedy.
L o B a t
L o B a t
L o B a t
L o B a t
L o B a t
Is indicated when batteries are weak. Approximately 20 minutes of operating time
remain.
Error 1
Error 1
Error 1
Error 1
Error 1
Indicates an overload condition.
Error 2
Error 2
Error 2
Error 2
Error 2
Indicates an underload condition.
Error 7
Error 7
Error 7
Error 7
Error 7
EEPROM data incorrect.
Error 14
Error 14
Error 14 Zero exceeds ZERO% and cannot be zeroed.
Error 14
Error 14
Err 21
Err 21
Err 21
Calibration data does not match current full scale, Grad and Cal Point settings. Settings
Err 21
Err 21
must be restored or the Indicator must be recalibrated using the current settings.

5.3 Service Information

If the Troubleshooting section does not resolve or describe your problem, you will need to contact an authorized
Ohaus Service Agent. For Service assistance in the United States, please call Aftermarket, Ohaus Corporation toll-
free at (800) 526-0659. An Ohaus Product Service Specialist will be available to help you.

5.4 Technical Data

Materials
Housing stainless-steel
Keypad/display overlay polyester

Ambient conditions

The technical data is valid under the following ambient conditions:
Ambient temperature
Relative humidity
Height above sea level
Operability is assured at ambient temperatures between 5 and 40° C.
-10°C to 40C/ 14°F to113°F
10%.......95%, noncondensing
up to 4000m