The customer or the user of the digital Blood Pressure Monitor type BW-BT1 can help
prevent electromagnetic interference by maintaining a minimum distance between port-
able and mobile RF communications equipment (transmitters) and the Digital Blood Pres-
sure Monitor type BW-BT1 as recommended below, according to the maximum output
power of the communications equipment.
output power
80 MHz to 80 MHz
of transmitter
d =
1.167 √
0.01
0.1
1
10
100
11.667
For transmitters rates at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmit-
ter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800MHz, the reparation distance for the higher frequency range
applies.
Note 2: these guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
102
BewellConnect - BW-BT1 - User Manual - 2018-03_V1
Separation distance according to
frequency of transmitter (m)
80 MHz to 2.5GHz
d =
2.333 √
P
P
0.117
0.233
0.369
0.738
1.167
2.333
3.689
7.379
23.333