DATOSPIR
· User's Manual
When "MEASUREMENT ERROR" visual indication is detected repeatedly,
switch off the device and contact SIBEL S.A.U. After Sales Service or your dealer.
It is recommended to change the battery before performing more spirometry maneuvers
when "LOW BATTERY" visual indication is shown. Previous maneuvers will not be lost since
they are stored in the W20s Spirometry Software.
3.5 TRANSPORT AND STORAGE
DATOSPIR
A carrying bag is included in all models of
that allows storing the device
and most of its accessories for and easy transportation. The device can be stored in this
bag while not in use. It is recommended to remove the batteries if the device will be not
used for a week or longer.
3.6 FIRMWARE UPDATE
To update the device's firmware to a new version connect the device to the PC via USB cable
and send the new firmware using the W20s software (see process description in the W20s
user's manual). When the update is complete, turn off the device and now it is ready for use.
If the device does not respond as expected, you can enter into the RECOVERY MODE by
holding down the key
for a few seconds until the visual indicator blinks rapidly in green.
Then, you can update the firmware using the W20s software.
In case that the communication is interrupted during the firmware update, the device will
enter into the BIOS MODE after restart. This mode allows updating again the firmware,
following the procedure described previously.
If you cannot update the firmware, contact
SIBEL
S.A.U. After Sales Service or your dealer.