1.
Remove the transducer by slightly
pressing the tab to release it from its
housing.
DO NOT rinse the turbine under the stream of tap water. The turbine can
be damaged.
2.
Immerse the transducer in a CIDEX® OPA disinfectant solution (see manufacturer's
instructions). Rinse with distilled water.
3.
Reliability depends on the condition of the turbine. Inspect it for possible damage.
4.
Dry it at room temperature and assemble again.
It is recommended to have several transducers to replace them while disinfecting the
transducers already used.
If an antibacterial filter is used with the Fleisch or Turbine transducers, the CIDEX® OPA
solution indicated in step 2 can be replaced with soapy water (neutral soap).
C) DISPOSABLE LILLY TRANSDUCER
DO NOT reuse Disposable transducer, disposable mouthpieces, disposable
filters or disposable spares for filters. Otherwise, there is risk of cross-
contamination between patients.
511-D00-MUM · Rev 1.01
DATOSPIR
· User's Manual
EN 19