Miscellaneous
EMC
first step select
mrs uses rf energy only for its internal function. therefore, its rf emissions are very low and are not likely to
™
cause interference in nearby electronic equipment. the first step select
• EN 60601-1-2: 2007
• Conducted Emissions: Nov. 21, 2008 (Group 1, Class B)
• Radiated Emissions: Dec. 17, 2008 (Group 1, Class B)
• EN 61000-4-2: Dec. 3, 2008 (6/8 kV)
• EN 61000-4-3: Dec. 1, 2008 (3 V/m)
• EN 61000-4-4: Dec. 18, 2008 (2 kV)
• EN 61000-4-5: Dec. 1, 2008 (1 kV)
• EN 61000-4-6: Dec. 24, 2008 (3 Vrms)
• EN 61000-4-8: Dec. 25, 2008 (3 A/m)
• EN 61000-4-11: Dec. 21, 2008 (>95%-0.5p, 30%-25p,>95%-250p)
• EN 61000-3-2: Dec. 21, 2008 (Class A)
• EN 61000-3-3: Dec. 24, 2008
Classification
Classification in accordance with UL 60601-1, CSA C22.2 No 601.1 and IEC. 60601-1
• Class II (UL 60601-1)
• type B
• no sterilization
• not for Use with flammable anesthetic mixture with air, oxygen or nitrous oxide.
• Continuous operation
• ordinary equipment
Environmental Storage / Transit Conditions
Temperature Range: -20˚ to 60˚ C (-4˚ to 140˚ F)
Relative Humidity Range: 10% to 95%
Atmospheric Pressure Range: 500 hPa to 1060 hPa
Environmental Operating Conditions
Temperature Range: 10˚ to 40˚ C (50˚ to 104˚ F)
Relative Humidity Range: 10% to 95%
Atmospheric Pressure Range: 500 hPa to 1060 hPa
mrs was tested to the following standards.
™
33