9.3 DC CURRENT
range
200 µA
2 mA
20 mA
200 mA
15 A
Overload protection: F0.5 A/600 V fuse, F15 A/600 V fuse
Continuous testing maximum steady-state current of 1 A
9.4 AC CURRENT
range
2 mA
20 mA
200 mA
15 A
Average sensing, calibrated to rms of sine wave
Frequency range: 40-500 Hz
Overload protection: F0.5 A/600 V fuse, F15 A/600 V fuse
Continuous testing maximum steady-state current of 1 A
9.5 RESISTANCE
range
200 Ω
2 kΩ
20 kΩ
200 kΩ
2 MΩ
V. 02 – 12/07/2018
DVM895
Do not measure circuits that may contain voltages
> 600 VDC or > 600 VAC
resolution
0.1 µA
1 µA
10 µA
100 µA
10 mA
Do not measure circuits that may contain voltages
> 600 VDC or > 600 VAC
resolution
1 µA
10 µA
100 µA
10 mA
Do not conduct resistance measurements on live
resolution
0.1 Ω
1 Ω
10 Ω
100 Ω
1 kΩ
accuracy
± (1.2 % rdg + 2 digits)
± (1.5 % rdg + 2 digits)
± (2.0 % rdg + 3 digits)
accuracy
± (1.5 % rdg + 3 digits)
± (2.0 % rdg + 3 digits)
± (2.5 % rdg + 5 digits)
circuits
accuracy
± (1.0 % rdg + 3 digits)
± (1.0 % rdg + 2 digits)