Error Messages - Beurer medical FT 90 Manual De Instrucciones

Termómetro clínico sin contacto
Ocultar thumbs Ver también para FT 90:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

If you want to use the thermometer to measure the room temperature you
need to switch to room temperature mode.
• With the thermometer switched on, briefl y press the MODE/MEM but-
ton twice.
The device switches to the room temperature mode
• The room temperature is immediately displayed (fi g. 10).
Measurements taken in room temperature mode are not stored.

9. Error messages

Error
Problem
message
Measurement during self-test,
device not yet ready for measure-
ment.
Large variation in room tempera-
ture.
Room temperature below <50 °F,
>104 °F (<10 °C or over 40 °C).
Faultless functionality cannot be
guaranteed.
.
Solution
Wait until the forehead symbol stops fl ash-
ing.
Store the device for at least 30 minutes
in a room with room temperature
(50 –104 °F / 10 – 40 °C ).
Room temperature must be between 50 °F
and 104 °F (10 °C and 40 °C).
Remove the battery for approx. 1 minute
and then reinsert. In the event of a
repeated error message, contact your
retailer or Customer Services.
Fig. 10
17

Capítulos

Tabla de contenido
loading

Tabla de contenido