Cuda surgical EX400 Manual De Usuario página 11

Iluminador de xenón con regulador de luminosidad digital
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 87
6.2.1.1
(Refer to 8.1 LAMP REPLACEMENT) Disconnect power. Open the door. The lamp
positioning lever should be completely down(horizontal) to properly engage the lamp
and completely up (vertical) to release it. Raise the lever completely if not already
raised and firmly push the lamp to assure full engagement to the power connectors.
Lower the lamp engagement lever counter-clockwise, from vertical to horizontal, and
observe that the lamp rises slightly as it is engaged into proper operating position.
6.2.1.2
6.2.1.3
The EX400 digital control system communicates with Cuda
for it. Only Cuda
®
LIT-238 CUDA
SURGICAL
(English)
ENGAGE LAMP
CLOSE DOOR
Close the door completely. (Fan will continue to run with door open.)
INCOMPATIBLE LAMP MODULE INSTALLED
Install correct SYC0450 lamp.
®
Surgical lamps can be engaged by the positioning mechanism.
®
Surgical lamps designed
Rev. -
Date of Revision: 06/08/18
Page 11 of 147
loading