USER GUIDE
CHAIR PS-251 + PA-260
06 OPERATION
TRACK ANCHOR SYSTEM
Use the track anchor system when you want to attach
a track to the chair to facilitate patient transport by
stairs.
INSTRUCTIONS FOR ANCHORING THE TRACK:
First, place the track in an extended position, reclining
on the tread bands as shown in figure 1.
With the chair unfolded, bring the lower hooks of the
chair closer to the track, so that they are aligned with
the track anchors, slightly lifting the chair attach the
lower anchors as shown in figure 2.
Then, turn the chair on the power track, with a dry blow,
you should hear a CLACK sound and the coupling will
have been produced correctly, see image 3.
To end, lift the assembly and close the track so that the
assembly is according to the image 4.
For the decoupling process, pull the handles located at
the top of the track and it will be released.
During the operation of the coupling and decoupling of the track, the chair shall remain at all
times without any patient and on a surface as flat possible.
If not, both patient and personnel may result in wounds and may result in damaged important
product elements.