Mythermo Bw-Cx10 Accuracy; Supplied Accessories; Troubleshooting - VISIOMED bewell connect MyThermo Manual De Usuario

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Accuracy
Bluetooth Technology
Power Consumption
Voltage
Power utilized in sleep mode
Power utilized when turned on
Automatic shutoff
DEVICE LONGEVITY: The MyThermo BW-CX10 is guaranteed for approximately 40,000 readings.

11. MYTHERMO BW-CX10 ACCURACY

From 30°C to 34.9°C = +/ - 0.4°C
From 86°F to 94.82°F =+/ - 31.3°F
From 35°C to 42°C = +/ -0.2°C
From 95°F to 215.96°F =+/-31.6°F
From 42.1°C to 43°C = +/ -0.4°C
From 107.78°F to 109.4°F =+/ - 31.3°F
Accuracy outside the measuring range (30°C to 43°C - 86°F to 109.4°F) cannot be guaranteed.

12. SUPPLIED ACCESSORIES

- User Manual
- Charging cable
- Base

13. TROUBLESHOOTING

If you have one of the following problems while using your MyThermo BW-CX10, please refer
to this troubleshooting service guide to help resolve the problem. If the problem persists,please
contact our customer service at http://www.support.bewell-connect.com.
16
+/-0.4°F / +/- 0.2°C in Body mode (95-107.6°F /
35°C-42°C)
+/- 1.8°F / +/- 1.0°C in Room mode
+/- 0.7 °F / +/- 0.4°C in Surface mode
4.0
< 50 mW
3.6V~4.2V
≤15μA
≤150mA
10~20 second
BewellConnect - BW-CX10 - User Manual - 042017v3
ISO 80601-2-56:2009
Tabla de contenido
loading

Este manual también es adecuado para:

Bewell connect bw-cx10

Tabla de contenido