Storz nova 150 Manual De Instrucciones página 47

Ocultar thumbs Ver también para nova 150:
Tabla de contenido
41
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
Table 9 – Test specifications for ENCLOSURE PORT IMMUNITY to
RF wireless communications equipment (according to IEC 60601-1-2)
Test frequency
Band
(MHz)
(MHz)
385
380 – 390
450
430 – 470
710
745
704 – 787
780
810
870
800 – 960
930
1 720
1 845
1 700 – 1 990
1 970
2 450
2 400 – 2 570
5 240
5 500
5 100 – 5 800
5 785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the LED nova 150 model 20 1612 20 may
be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
Electromagnetic
Compatibility
(EMC) Information
Service
Modulation
Pulse modulation
TETRA 400
18 Hz
FM
GMRS 460,
± 5 kHz deviation
FRS 460
1 kHz sine
Pulse modulation
LTE Band 13, 17
217 Hz
GSM 800/900,
TETRA 800,
Pulse modulation
iDEN 820,
18 Hz
CDMA 850,
LTE Band 5
GSM 1800;
CDMA 1900;
GSM 1900;
Pulse modulation
DECT;
217 Hz
LTE Band 1, 3,
4, 25; UMTS
Bluetooth,
WLAN 802.11 b/g/n,
Pulse modulation
RFID 2450,
217 Hz
LTE Band 7
Pulse modulation
WLAN 802.11 a/n
217 Hz
Indicaciones sobre
compatibilidad
electromagnética (CEM)
Maximum
Distance
Power
(W)
(m)
1.8
0.3
2
0.3
0.2
0.3
2
0.3
2
0.3
2
0.3
0.2
0.3
IMMUNITY
TEST LEVEL
(V/m)
27
28
9
28
28
28
9
Tabla de contenido
loading

Tabla de contenido