the correct positioning of the grounding plate;
•
the correct connection of the cables and their connectors;
•
the correct activation of the starting keys (hand-switch of foot-control);
•
that there is no damage to the insulation of the cables;
•
that the electrode is not dirty.
POSSIBLE CAUSES OF BURNS
Burns may be caused by a high density of current in the patient's tissues or by the heating of
inflammable fluids or gases. Possible causes are:
•
The patient has inadvertently been placed in contact with electrically-conductive parts.
•
A direct contact has occurred between the cables and the patient's skin, which has caused a
capacitive effect.
•
The combustion of inflammable disinfectant agents.
•
The combustion of inflammable narcotic gases.
CONTRA-INDICATIONS
The products must not be utilised if:
•
there is visible damage to the electrode or extension, to its insulation, to the body and/or
connector of the pencil, or to the connection cable;
•
the preliminary operational check gave negative results;
•
the electrode or extension is not firmly inserted in the distal part of the product.
STORAGE
The package must be stored at temperatures between 0°C and 50°C and at a relative humidity of
between 20% and 80%.
GUARANTEE AND RESTRICTIONS
FIAB guarantees that the product complies with the requirements of Directive 93/42/EEC. The
manufacturer cannot be held responsible, nor be made liable, for refunding medical expenses for direct
or indirect damages deriving from the non-functioning or anomalies of the models described above,
should the products not have been utilised according to these instructions for use. It is advisable to
inform the FIAB Quality Assurance Service regarding any type of malfunction or defects that may come
to light with regard to this device.
DISPOSAL
Waste matter coming from hospitals must be disposed of in compliance with existing regulations.
MEHRWEGSTIFTE, ELEKTRODEN UND
DE
MEHRWEGVERLÄNGERUNGEN FÜR ELEKTROCHIRURGIE
GEBRAUCHSANWEISUNG
ANWENDUNG:
Schnitt und Gerinnung von Geweben während elektrochirurgischer Verfahren unter Verwendung eines
kompatiblen Hochfrequenzgenerators.
MEHRWEGSTIFTE MIT MANUELLER BEDIENUNG, ANSCHLUSS TYP VALLEYLAB
ELEKTRODENMODELLE FÜR 2,38mm (3/32") - ANSCHLUSS: F4141 - F4242 - F4243 - F4244 (mit
Standardklinge), F4141/AGO (Nadelelektrode), F4151 - F4252 - F4253 - F4254 (mit 5m langem Kabel
und Standardklinge).
F4244/S (steril).
ELEKTRODENMODELLE FÜR 4mm - ANSCHLUSS: F4246 (Klingenelektrode),
langem Kabel und Klingenelektrode).
MEHRWEGSTIFTE MIT FUSSBEDIENUNG, ANSCHLUSS PIN 4mm
MODELL: F4812 - F4814 (mit Standardklinge).
9 / 32
F4256 (mit 5m