Symbol
Imbalance: centrifuge stops
Erro1
due to a detection of an
excess of vibrations.
Over speed: centrifuge stops
Erro2
due to a detection of over
speed of motor.
Lid
protection
Erro3
acting.
Erro4
Input voltaje too low
Erro5
Braking over voltage
Erro6
Over current
Erro7
No speed measurement
Fault
of
Erro8
between internal components
of the equipment.
Erro9
Over voltage
Chart 3: Error symbols and their causes
Cause
● Re-balance the samples; the maximum error allowed is
● The device is placed in a leaning position, not horizontal;
● The motor shaft is bent; equipment needs to be
repaired by Technical Service Department.
● Problem of microcomputer control system; equipment
● Problem of speed sensor; equipment needs to be repaired
by Technical Service Department.
● Re-close the lid
system
is
● Lid protection system is broken; equipment needs to be
repaired by Technical Service Department.
● Check the power source
● The Power source voltage is too low or the braking
resistor failed or brake too fast; equipment needs to be
repaired by Technical Service Department.
● The Power source voltage is too low or the
acceleration is too fast; equipment needs to be
repaired by Technical Service Department.
● Speed sensor or motor failed; equipment needs to be
repaired by Technical Service Department.
communication
● Equipment needs to be repaired by Technical
Service Department.
● The Power source voltage is too low or the
acceleration is too fast; equipment needs to be
repaired by Technical Service Department.
≤1.5g
change position of equipment so as it is perfectly
horizontal.
needs to be repair by Technical Service Department.