Active from setting
This is a percentage of the max. nominal value of the device at which the click or dial
mode starts to record peak values.
Setpoint value display
When a target value is selected, the torque tester indicates in three different ways
whether the reading is too low, within tolerance or too high.
1.
Text colour: low = yellow; within tolerance = green; high = red
2.
Setpoint value symbols: low = ; within tolerance = ü; high =
3.
Setpoint value display line
Example
Setpoint value 1 (T01) set to 200.0 Nm ± 3 % (194.0 - 206.0 Nm); mode set to click
First measurement: 190.7 Nm (value too low)
Clockwise
Low
Below tolerance
Second measurement: 202.1 Nm (value within tolerance)
Within
tolerance
Within
tolerance
Third measurement: 212.4 Nm (value too high)
High
Above
tolerance
26
Operating instructions
2
The setpoint value display
line is scaled to show
± 20 % in relation to the
setpoint value (200 Nm)
The two lines directly next
to the centre line show the
tolerance range for the
setpoint values (± 3%)